Samsung provides reliable cleaning guidelines to ensure confidence in equipment safety.
Our expert-approved protocols deliver clear, step-by-step instructions for thorough cleaning and disinfection, reinforcing your commitment to patient care.
Search by
Categories
Cleaner, disinfectants, and gel option (up to 5 items)
Custom filter
Transducer
X
= Not compatible (DO NOT USE)
Blank
= Untested (DO NOT USE)
○
= Compatible (From Lens to Housing)
●
= Compatible (From Lens up to 50cm away from the connector)
○(0*)
= Must not be used longer than (*5/10/15/20/25/30/50) minutes.
★
= Staining may occur on surface of transducer; however, the acoustic performance and image quality are not affected.
◆
= Compatible and Validated for disinfection method
◆/★
= Compatible and Validated but staining may occur on the exterior; however, the acoustic performance and image quality are not affected.
All transducers are under tested in IPx 7 Criteria.
Be sure to check the manufacturer's instructions for the dilution strength and method for using powders including Perasafe.
V-PRO products include V-PRO1, V-PRO1 Plus, V-PRO Max, V-PRO Max2, V-PRO 60 sterilizer.
- For guaranteed performance, this device has been tested and confirmed for up to 200 cycles.
Be sure to follow the manufacturer's instructions before using V-pro sterilization.
Be sure to check the manufacturer's instructions before using Antigermix disinfection system to prevent discoloration.
Do not use ultrasound gels or couplant that contain any of the following ingredients:
- Oil-based lotion type couplants, Mineral oil ,Olive oil, Lubricant oil, Paraffin-based grease, Dimethylsilicone, Alcohol such as methanol,
- Ethanol or isopropanol, Acetone ,Dioctylphtalate, Aromatic substances, Iodine compounds, Aloe Vera
Non-recommended Gel - POLYSONIC® ULTRASOUND LOTION
The compatibility indicated by the symbols (Legend) in this table represents material compatibility only, based on physical and mechanical testing. The provided Disinfectant Level information is for reference purposes only.
Therefore, users must not rely solely on the compatibility symbols or the disinfectant levels indicated in this table. It is mandatory to consult the disinfectant manufacturer’s official Instructions for Use (IFU) to ensure compliance with the required disinfection level, clinical application, and local regulations.